Abstract
We explored whether a noninvasive handheld device using Active Cancellation of Tremor (ACT) technology could stabilize tremor-induced motion of a spoon in individuals with essential tremor (ET). Fifteen ET subjects (9 men, 6 women) performed 3 tasks with the ACT device turned on and off. Tremor severity was rated with the Fahn-Tolosa-Marin Tremor Rating Scale (TRS). Subjective improvement was rated by subjects with the Clinical Global Impression Scale (CGI-S). Tremor amplitude was measured using device-embedded accelerometers in 11 subjects. TRS scores improved with ACT on (versus off) in all 3 tasks: holding (1.00 ± 0.76 vs. 0.27 ± 0.70; P = 0.016), eating (1.47 ± 1.06 vs. 0.13 ± 0.64; P = 0.001), and transferring (1.33 ± 0.82 vs. 0.27 ± 0.59; P = 0.001). CGI-S improved with eating and transferring, but not the holding task. Accelerometer measurements demonstrated 71% to 76% reduction in tremor with the ACT device on. This noninvasive handheld ACT device can reduce tremor amplitude and severity for eating and transferring tasks in individuals with ET.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.