The authors are with the Applied Physiology Branch. Environmental Science Division. U.S. Air Force School of Aerospace Medicine, Brooks AFB, Tex. 78235. A nondestructive technique to continuously monitor diameter and blood flow at the same point along the descending thoracic aorta has been developed. A probe was passed down the esophagus to a level 1 to 2 cm distal to the aortic arch. On its distal end were Doppler-shift velocity-detecting and pulse-echo position- sensing crystals. The pulse-echo system was used not only to measure the diameter of the aorta, but also to aim all of the probe crystals directly across the axis of the aorta. Information supplied by this system aided in the identification of the aorta and helped to avoid measurement artifacts such as the inclusion of other arteries, veins, or moving structures.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.