Abstract

A non-derivative method was developed for the qualitative and quantitative analysis of isosteroidal alkaloids from Fritillaria thunbergii. During method development the performance of two universal detectors, the charged aerosol detector (CAD) and evaporative light scattering detector (ELSD), were evaluated. The CAD was found to be 30 to 55 times more sensitive than ELSD enabling the measurement of low levels of reference compound impurities that could not be detected by ELSD. The peak area percent of the reference compound, peimisine, obtained by CAD was 50.10%, but 91.66% by ELSD showing that CAD is suitable to estimate the presence of impurities. The CAD showed good reproducibility with overall intra- and inter-day peak area RSD values of less than 1.8% and 2.7%, respectively and had a linear dynamic range of up to 4 orders of magnitude (0.06–44mg/L) for peimine and peiminine. The optimized method was used for the quantitative analysis of peimine and peiminine from F. thunbergii.