Abstract
We have developed a miniature fiber-optic probe with no focusing optics for in situ analysis of volatile organic compounds (VOCs). The probe uses an optical fiber to transmit a laser pulse to a vapor sample causing it to ionize adjacent to the fiber tip through a resonance-enhanced multiphoton ionization (REMPI) process. The distal end of the optical fiber is contained co-axially within 2-mm-inner-diameter stainless steel tubing that serves as an electrode. The electrode is biased at a high positive potential to collect electrons. The current generated is shown to be proportional over about two orders of magnitude to the concentration of the species ionized. Visible wavelength REMPI spectroscopy is used to determine probe sensitivities of 20 ppb (benzene) and 43 ppb (toluene). Designing the probe without focusing optics specifies an achromatic ionization region constant in size and position as the laser wavelength is scanned, which simplifies data collection and reduction. Focusing achromatic systems are discussed and the potential signal improvement is estimated.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.