Abstract

From the viewpoint of plasma particle measurements in the radiation belt, background noise is a serious problem. High-energy particles penetrating the sensor shielding generate spurious signals, and their count rate often can be comparable to the true signals. In order to attenuate such background noise during medium-energy (5–83 keV) electron measurements, we propose the double energy analyses (DEA) method. DEA is conducted by a combination of an electrostatic analyser (ESA) and avalanche photo-diodes (APDs); ESA and APD independently determine the energy of each incoming particle. By using the DEA method, therefore, the penetrating particles can be rejected when the two energy determinations are inconsistent; spurious noise are caused only when the deposited energy at an APD is by chance consistent with the measured energy by ESA. We formulate the noise count rate and show the advantage of DEA method quantitatively.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.