Abstract

Microcalcifications within the fibrous cap of the arteriosclerotic plaques lead to the accrual of plaque-destabilizing mechanical stress. New techniques for plaque screening with small detectors and the ability to differentiate between the smooth and hard elements of plaque formation are necessary. Vascular plaque formations are characterized as calcium phosphate containing structures organized as hydroxylapatite resembling the mineral whitlockite. In transmission and reflexion studies with a simple millimeter wave (mm-wave)-demonstrator, we found that there is a narrow window for plaque detection in arterial vessels because of the tissue water content, the differentiation to fatty tissue, and the dielectric property of air or water, respectively. The new sensor is based on a sensing oscillator working around 27 GHz. The open-stub capacitance determines the operating frequency of the sensor oscillator. The capacitance depends on the dielectric properties of the surrounding material. The sensor components were completely built up in surface mount technique. Completed with a catheter, the sensor based on microwave technology appears as a robust tool ready for further clinical use.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.