Abstract
For pt.I see ibid., vol.38, p.1-16, Jan. 1991. The signal models and performance of the estimation strategies described in pt.I are tested with experimental ultrasonic data. The ultrasonic data analyzed verify the theoretical model and predicted performance. The averaged correlation, verified experimentally, confirms that the correlation envelope can be used to estimate the velocity of scatterers and that the shape of the correlation function conveys information regarding the velocity profile within the sample volume. For both the wideband point and range spread estimators, the predicted improvement in velocity resolution and the reduction in height of subsidiary velocity peaks are demonstrated. Through the use of these estimation strategies, information regarding the mean velocity and velocity variation are available for each spatial location within the vessel. This information is presented using a three-dimensional spatial velocity profile display, which appears to offer a number of advantages in the rapid identification of pathology.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.