Abstract
Two-fold astigmatism, three-fold astigmatism and misalignment (coma) can be found from the orientations only of the diffractograms of a set of images with injected beam tilts – typically eight images with tilts of equal magnitude and azimuths 45° apart. The spherical aberration must be determined independently, and the defocus is not measured; however, the measurement required is very much simpler than the measurement of image displacements or induced defocus/astigmatism values, and no calibration of the microscope magnification is needed.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.