Abstract

A new ultrasonic root canal preparation system has been developed that electronically monitors the location of the file tip during all instrumentation procedures. The Root ZX has been adapted for this purpose, and its filter circuit effectively cuts out the large spike noise of the ultrasonic unit. During enlargement of the canal, the ultrasonic vibration of the file can be stopped at any desired position on the meter. Extracted human tooth models with electronically measurable canals were used to test the device. Pre- and postoperative shapes of the root canals were evaluated using contact microradiography. The autostop mechanism worked correctly. Using a weak power and fine files, straightening, ledge formation, and file breakage were minimal. It seems that this system minimized the danger of overinstrumentation and could be safely applied in clinical practice.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.