Abstract

Augmentation index (AIx) and subendocardial viability ratio (SEVR) are widely accepted indices of wave reflection and myocardial oxygen demand relative to supply. This study aimed to validate a new tonometric device (IIM-2010A) for obtaining AIx and SEVR from radial artery. A total of 68 outpatients (32 men and 36 women) aged 20 to 76 years (44.7±16.6 years) recruited from a health screening center participated in the study. AIx was obtained from radial pressure using the HEM-9000AI and IIM-2010A devices, while SEVR was measured from carotid pressure with the tonometric method and from radial pressure by the IIM-2010A device. In a subgroup of 24 patients, the measurements of AIx and SEVR were repeated after an interval of 10 minutes. The correlation of radial AIx between the IIM-2010A and HEM-9000AI devices was highly significant (r=0.956, P<.01). Radial SEVR determined from IIM-2010A was also highly related to carotid SEVR (r=0.864, P<.01), although the value was about 13.1% lower. There was no statistically significant difference between the repeated measurements of both indices. The lower coefficient of variation (2.9% vs 4.3% for AIx, 3.3% vs 4.1% for SEVR) and higher intraclass correlation coefficient (0.96 vs 0.91 for AIx, 0.93 vs 0.86 for SEVR) of IIM-2010A confirmed better short-term reproducibility, compared with the HEM-9000AI device and carotid tonometry. The new tonometric device IIM-2010A is effective and reproducible in calculating radial AIx and SEVR and has potential use in health screening.