Abstract

To overcome the temperature effect of NaI(Tl) detectors for energy spectrometry without additional hardware, a new correction method was put forward based on pulse deconvolution, trapezoidal shaping and amplitude correction, named DTSAC. To verify this method, actual pulses from a NaI(Tl)-PMT detector were measured at various temperatures from -20 °C to 50 °C. Pulse processing and spectrum synthesis showed that the position drift of the 137Cs 662 keV peak was less than 3 keV, and the corresponding resolution at 662 keV of the sum spectra ranged from 6.91% to 10.60% with the trapezoidal width set from 1000 ns to 100 ns. The DTSAC method corrects the temperature effect via pulse processing, and needs no reference peak, reference spectrum or additional circuits. The method solves the problem of correction of pulse shape and pulse amplitude at the same time, and can be used even at a high counting rate.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.