Abstract

Bio-impedance spectroscopy (BIS) is a sophisticated testing technique used to analyze impedance changes at different frequencies. In this study, we investigated the estimation of the Cole Model for BIS measurements without the need for high-frequency resistance and reactance measurements, where they are inaccurate due to leakage capacitences. We employed a Texas Instruments evaluation kit (AFE4300) and compared the Cole plots of two different circuit models of tissue between the proposed configuration and a commercial impedance analyzer used as a reference. To enhance the performance of the AFE4300, we incorporated an external direct digital synthesis (DDS) to generate higher frequencies. The results demonstrated the reliability of the proposed theoretical estimation technique in accurately estimating the resistances and capacitance of the Cole Model.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.