Abstract

AbstractA new technique is presented for the determination of concentration depth profiles. Surface atoms are sputtered by an ion beam and deposited on a clean silicon wafer. The wafer is rotated behind a slit in step with the sputtering progress. In this way the depth profile of the sample is transferred into a lateral distribution of the sputtered atoms on the target wafer. Subsequently the wafer is scanned by Total Reflection X-ray Fluorescence Spectrometry (TXRF) which is capable of detecting traces of metallic impurities on wafers down to 10-4 of an atomic monolayer. The sequence of ion-beam sputtering, deposition of the sputtered atoms and TXRF analysis results in an excellent depth resolution in the case of areal structures. Using an ion source of the Kaufmann type, an extrapolated perpendicular resolution better than 0.1 nm was obtained for a 1500 mm2 surface. For a surface area of 3 mm2 a depth resolution of 1 nm is expected. 1.4 nm was actually measured to be the width of a coherent Ti/Al-interface within a layered structure.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.