Abstract

Recording electrical auditory brainstem responses (EABR) provides clinical insight about responses of the residual post-cochlear neural system to electrical stimulation in profoundly deaf patients. A new strategy is presented for stimulating patients already implanted with a 15-electrode cochlear implant. Since the device is fully re-programmable via a RS-232 PC interface, it was possible to load a specific stimulating strategy designed to improve the spatial locus and the temporal structure of the impulse stimulation. Waves III to V emerge more clearly when this method is applied.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.