For many years, the irradiation of small volumes of tissue in the brain to necrotizing doses has been investigated as a “non-invasive” alternative to neurosurgery. We propose a new system in which a precisely machined helmet serves as a multi-port focussed X-ray collimator when it is itself irradiated by a conventional medical linear accelerator run in the electron mode. When the collimator is attached to a stereotactic frame, the geometric accuracy of delivering small radiation fields to the brain is limited primarily by the accuracy of the stereotactic localization, and is relatively independent of the positional stability of the accelerator. Field sizes as small as two millimeters are readily achievable. The problem of low dose rate associated with these small fields is overcome by the use of high electron beam currents.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: International Journal of Radiation Oncology, Biology, Physics
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.