Abstract
A method using a conventional retarding potential difference ion source in a new way to study charge transfer reactions is described. Potentials on five focus plates between the filament and ion box are applied so as to accelerate electrons, produce positive ions, decelerate the electrons and accelerate the positive ion beam. Only the primary positive ion beam traverses the ion box and is collected on the trap. This primary ion beam has momentum perpendicular to the mass analysis slit and is not collected. If charge transfer from the primary ions to a gas in the ion box occurs, the secondary ions are collected through the analysis slit with high efficiency. This method has features in common with the Cermak—Herman mode of ion source operation but also provides measurement of the primary ion beam. Thus dissociative charge transfer and total charge transfer between different molecules can be measured. Results compre favorably with those obtained by tandem mass spectrometers.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: International Journal of Mass Spectrometry and Ion Physics
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.