Abstract
A new bio-tribological simulator system was designed and built for basic wear tests on artificial hip joint materials and for common tribological studies applications. The module of hip joint simulator is consisted of Flexion and Extension (FE), Abduction and Adduction (AA) and Internal and External Rotation (IER). Preliminary tests were done with a 28 mm CoCrMo femoral head and a Ultra High Molecular Weight Polyethylene (UHMWPE) cup. Results showed that the wear rate was close to the clinical one. A frequency control system and a heating system were developed to offer a wide range of rotation frequency and temperature so as to provide proper experiments. In the Pin-on-Disk (POD) part, eight precision-made pins were generated and an advanced computer system was built up to measure the friction coefficient of the tests samples.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.