Abstract

The disposable device for intraoperative reinfusion consists of a polycarbonate container with an antifoam-coated filter. One port is connected to the suction tip; the other port is attached to wall suction. Sodium citrate solution (50 to 100 ml) is aspirated first, followed by 400 ml of shed blood. The full container is passed to the anesthetist who turns the device upside down and starts reinfusion from the bottom, previously the top port of the container. In experiments with dogs, this procedure resulted in only slight hemolysis, a minor drop in platelet count, and no mortality or morbidity. This new method of reinfusion combines the advantages of simplicity, disposable equipment, and low blood trauma.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.