Abstract
The recently developed technique of fibre optic respiratory plethysmography (FORP) has been modified to monitor the rapid, small amplitude movements of the chest wall during high-frequency oscillatory ventilation (HFOV). The FORP sensor is an expandable belt encircling the chest, in which is housed a fibre optic loop that alters its radius of curvature as a function of chest perimeter. These curvature changes cause variations in macrobending losses of light transmitted through the fibre, which are proportional to the chest perimeter. Dynamic measurement of transmitted light intensity can hence be used to monitor chest wall motion (CWM). For application to HFOV, the design of the FORP belt was altered to increase sensitivity and the materials were chosen to maximise macrobending effects induced by the CWM. FORP was tested in four piglets ventilated with HFOV, both in the normal and surfactant-deficient lung. Measurement of CWM was possible over the full range of tidal volumes and ventilation frequencies used during HFOV. In all cases, the measured frequency of the CWM fell within 3% of the applied ventilation frequency. In addition, the technique was sufficiently sensitive to detect changes in the amplitude of CWM in response to changes in applied tidal volume. It is anticipated that application of this new non-invasive measurement device will lead to an increased understanding of the dynamics of chest and abdominal wall motion during HFOV.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.