Abstract
Secondary ion mass spectrometry (SIMS) with gas cluster ion beams (GCIB) was investigated with experiments and molecular dynamics (MD) simulations. A new cluster SIMS system with high-intensity cluster ion source has been developed. Ar cluster ion beam with the average size of 2000 was utilized as a primary ion beam. The beam diameter is less than 1 mm at the target position, and a current density of 10 μA/cm 2 is obtained. The etch rate of Si with this current density is more than 20 nm/min, which is far beyond the etch rate in recent low energy SIMS system. The secondary ion intensity linearly increases with the acceleration voltage. A threshold voltage of a few keV for secondary ion emission was found. These results are consistent with previous sputtering experiments.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Nuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.