Abstract

We investigated the accuracy and safety of a new HURWA robotic-assisted total knee arthroplasty (TKA) system in a sheep model. Ten male small-tailed Han sheep were used in this study. Sheep were imaged by computed tomography scan before and after bone resection and the cutting errors between actual bone preparation and preoperative planning of the femur and tibia in three dimensions were measured. The overall accuracies after surgery compared with that from preoperative surgical planning of the left and right femurs were 1.93±1.02° and 1.93±1.23°, respectively. Additionally, similarly high overall accuracies for the left and right tibia of 1.26±1.04 and 1.68±0.92°, respectively, were obtained. The gap distances of the distal cut, anterior chamfer, anterior cut, posterior chamfer and posterior cut on the medial side were 0.47±0.35mm, 0.41±0.37mm, 0.12±0.26mm, 0.41±0.44mm and 0.12±0.23mm, respectively. No intraoperative complications, such as intraoperative fracture, massive bleeding or death, occurred. This new HURWA robotic-assisted TKA system is an accurate and safe tool for TKA surgery based on the sheep model.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.