BackgroundEver since the introduction of laparoscopic surgery, researchers have been trying to add steerability to instruments to allow the surgeon to operate with better reachability and less tissue interaction force. Traditional solutions to introduce this often use a combination of springs, cables, pulleys, and guiding structures, resulting in instruments that cannot be properly cleaned and thus are very costly to manufacture and maintain. The aim of the study is to develop a novel affordable, sustainable, cableless, and fully steerable laparoscopic grasper, and to test its ease of assembly, disassembly, and use.MethodsA set of requirements was defined to ensure that the instrument can be handled efficiently at the sterilization unit and in the operating room. Based on these, a multisteerable, cableless 5 mm laparoscopic instrument that operates based on shaft rotations was developed. To test its assembly and disassembly, ten participants were asked to fully dismantle the instrument and reassemble it a total of 60 times. In addition, ten medical students were asked to use the grasper in the ForceSense box-trainer system on a newly developed 3D pick-and-place task, to determine the control effort based on learning curves of steering errors, task time, instrument path length, and maximum tissue interaction force.ResultsAll important design requirements were met. The recorded data indicates that ten engineering students were able to fully dismantle and reassemble the instrument shaft in 12 s (SD7) and 65 s (SD43) seconds at the sixth attempt. The learning-curve data indicates that three attempts were needed before the ten medical students started to use all steering functions. At the sixth attempt, on average only 1.25 (SD0.7) steering errors were made. The steepest slope in the learning curves for steering errors, path length, and task time was experienced during the first three attempts. In respect of the interaction force, no learning effect was observed.ConclusionThe multi-DOF (degree of freedom) cableless grasper can be assembled and disassembled for cleaning and sterilization within an acceptable time frame. The handle interface proved to be intuitive enough for novices to conduct a complex 3D pick-and-place task in a training setting.

Highlights

  • Ever since the introduction of laparoscopic surgery, researchers have been trying to add steerability to instruments to allow the surgeon to operate with better reachability and less tissue interaction force

  • Unlike the hinges in currently available instruments such as EndoWrist, AutoSuture, FlexDex and Endo Grasp, the shaft-actuated tip articulation (SATA) mechanism can be disassembled for cleaning and has a smaller bend radius, making it ideal for endoscopic procedures performed in small spaces like joints or hollow organs

  • As the tolerances between it and the tube are less than 0.1 mm, gas leakage through the instrument shaft is negligible and so pneumoperitoneum is maintained

Read more

Summary

Introduction

Ever since the introduction of laparoscopic surgery, researchers have been trying to add steerability to instruments to allow the surgeon to operate with better reachability and less tissue interaction force. The aim of the study is to develop a novel affordable, sustainable, cableless, and fully steerable laparoscopic grasper, and to test its ease of assembly, disassembly, and use. Ten medical students were asked to use the grasper in the ForceSense box-trainer system on a newly developed 3D pick-and-place task, to determine the control effort based on learning curves of steering errors, task time, instrument path length, and maximum tissue interaction force. The recorded data indicates that ten engineering students were able to fully dismantle and reassemble the instrument shaft in 12 s (SD7) and 65 s (SD43) seconds at the sixth attempt. The steepest slope in the learning curves for steering errors, path length, and task time was experienced during the first three attempts. The handle interface proved to be intuitive enough for novices to conduct a complex 3D pick-and-place task in a training setting

Objectives
Methods
Results
Discussion
Conclusion
Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.