Abstract
Despite the obvious advantages of cryogen-free magnets for NMR such as independence of liquid helium supply and the possibility to use the same magnet at different fields, the practical application of those magnets remains limited because of temporal magnetic field distortions associated with cryogen-free cold head operation. A new experimental method for the simple and reliable detection of the temporal field distortions is described in this paper. The accuracy of the magnetic field measurements by this method is two orders of magnitude higher than by conventional MetroLab Tesla meter. This has enabled us to make improvements in the design of cryogen-free magnets by reducing the amplitude of such field distortions down to sub ppb level. This then results in cryogen-free magnets that are suitable for MRI and MAS NMR applications.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.