Abstract

This method has been designed to determine the levels of CoQ10 in human blood which utilizes only one drop of blood (0.05 ml). This method encourages and facilitates more frequent monitoring of blood levels of CoQ10, and of clinical importance, reveals the presence or absence of compliance. CoQ11 was used as an internal standard. The mean recovery of CoQ10 was 96%. The correlation coefficient between using 1.0 ml and 0.1 ml of blood and between using 1.0 and 0.05 ml was 0.997 in the two comparisons. Therefore, it can be accurate to monitor the level of CoQ10 in a-"finger-prick"-drop of blood, and at least 24 such determinations can be conducted in a 2-day period.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.