Abstract
The paper describes a new method for time interval measurement for high resolution nuclear timing spectroscopy system. Unlike the conventional method described in this paper, the method makes use of high resolution (16 bit ADC) with low conversion time of nearly 1 μs along with a linear ramp which is generated at the instant of START pulse. The ramp output is sampled at the instant of STOP pulse and digitized with this ADC to measure the time interval with resolution of 16 bit. The system has been constructed using micro-controller based PC interface circuit and results have been reported in this paper. The system offers high resolution of 390 ps with wide range of 25.6 μs.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.