Abstract

A prospective clinical study of 50 patients was conducted to validate a new method of imageless computer navigated hip arthroplasty. The new method enables the surgeon to acquire all registration points with the patient positioned and draped in lateral decubitus position. The final component orientation was measured from post-operative CT scans. The mean error in component position was −1.1° (SD 3.1°) for inclination and 0.9° (SD 4.3°) for anteversion. This compared favourably with the error of −1.8° (SD 1.8°) for inclination and −4.8° (SD 2.7°) for anteversion when using the traditional APP registration. Results show that one can expect the acetabular component to be within a safe zone of ±10° in 99.8% for inclination and 97.7% for anteversion when using the new lateral registration method. Level of Evidence Level II, Prognostic study.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.