Abstract
Capillary zone electrophoresis was employed for the determination of pipemidic acid using an end-column amperometric detection with a carbon fiber microdisk array electrode, at a constant potential of -1.10 V vs. saturated calomel electrode. The optimum conditions of separation and detection were 1.2 x 10(-4) mol/LNaOAc - 8.8 x 10(-4) mol/ LHOAc for the buffer solution, 20 kV for the separation voltage, 5 kV and 10 s for the injection voltage and the injection time. The limit of detection was 1.05 x 10(-7) mol/L or 189 amol (S/N=3). The relative standard deviation was 0.31% for the migration time and 2.0% for the electrophoretic peak current. The method was applied to determining pipemidic acid in human serum.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.