Abstract

Blood pressure reflects a person's health.It is proposed here that the method of detecting blood pressure may be the key to improving the precision of blood pressure measurements. The oscillometric blood pressure measurement technique is widely used in automatic blood pressure measurement instruments correctly. A method of blood pressure measurement by oscillometric method is first presented. In the oscillometric method, the basic principle of the "feature point" method and the "amplitude characteristic ratios" method is also explained and discussed here. A new method of blood pressure measurement, namely the coefficient difference comparative method, is proposed here,which is based on the feature point method and amplitude characteristic ratios method. The method is proved both effective and reliable through the analysis of many cases and clinical tests. Utilizing Visual C++, software for this new and novel method was developed and passed criterion simulation apparatus test. When applied in hospital situation, its error was ±5%. It is concluded that the oscillometric blood pressure measurement method can provide better means of blood pressure measurements reference for doctors.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.