Abstract

A new method is described of the Zeeman effect atomic absorption analysis by means of observation in the direction transverse to a magnetic field without using any isotope. When a light source is placed in a suitable magnetic field and observed in the direction transverse to it, it is possible to make the π components coincide with an atomic absorption line and to displace the σ components from it. The π components are used as the probe light for the measurement of absorption and the σ components as the reference light for the correction of scattering and molecular absorption. Though the mercury resonance line at 2536.5 Å has complex hyperfine structure, this method permits the use of a lamp of naturally occurring mercury. Background absorption can be corrected accurately with our method, allowing analyses of mercury contents in various samples without any chemical pretreatment. This method is generally applicable also to accurate and highly sensitive analyses of elements other than mercury. A high-resolution spectrophotometer with magnetic scanning was used to observe atomic absorption profiles under various conditions.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.