Abstract
Purpose An objective, quantitative, and sensitive method to map retinal thickness is needed to diagnose more effectively the conditions causing alterations in thickness, such as macular edema and neuroretinal atrophy. Methods An instrument, the retinal thickness analyzer, was developed into a rapid scanning instrument, capable of covering macular areas of 2 x 2 mm in 200 or 400 msec and generating a detailed map of the retinal thickness. The performance was assessed in vitro and in five normal subjects who were scanned on three separate visits. Results Optimal depth precision was 5 to 10 microns, and the optimal depth resolution was 50 microns. Reproducibility was +/- 12 microns on the same day, +/- 13 microns for single maps obtained in multiple visits, and +/- 10 microns for three averaged maps per visit obtained in multiple visits. Conclusions This new method to analyze retinal thickness provides four unique features: multiple optical cross-sectioning of the retina, mapping of retinal thickness, high reproducibility, and short acquisition time. These capabilities promise to improve the diagnosis and management of common diseases such as macular edema and glaucoma.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.