A new method to correct the barrel distortion of an electronic endoscope image is presented. A correction model assuming circularly symmetric distortion is introduced with the following model parameters: the center of distortion and the coefficients of polynomials representing the distortion correction in the radial direction. If the imaging system is distortion-free, straight lines in the object space should be imaged as straight lines. Based on this criterion, a distorted image of a standard pattern consisting of a grid of several straight lines is recorded, and the model parameters are then estimated as a basis for straightening distorted lines. This method has the advantage of not needing a careful placement of the standard pattern for calibration. Correction results are presented for the grid pattern chart to verify a sufficient degree of correction. Examples of distortion correction of real intestinal images and physicians' comments are also presented.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.