Abstract

A new analytical determination method of homocystine in human plasma has been developed. The method utilises liquid chromatography coupled to ionspray tandem mass spectrometry. Quantitative analysis was achieved using as an internal standard homocystine-d8. Mass spectrometer operated in the multiple reaction mode: homocystine and homocystine-d8 were detected through the transition from the precursor to the product ion (from m/z 269.3 to 90.0, and m/z 277.3 to 94.0, respectively). The method is extremely sensitive, with limit of detection in the range of 6 fmol/L. The interassay and intraassay coefficients of variation for homocystine were 6.22% and 3.4%, respectively. The accuracy for the added homocystine ranged from 85% to 110%. High specificity of tandem mass spectrometry coupled with a fast chromatographic process is suitable for a rapid and reliable assay of homocystine.