Abstract

In this study, a new method was proposed for calibrating hygrometers. The moisture permeation cylinder, which was used to measure the water vapor permeability of fabrics, was used to generate various levels of relative humidity. A dry nitrogen stream was employed to sweep the water vapor away. It was deduced that the relative humidity of the outgoing nitrogen stream was negatively correlated to the volumetric flow rate of nitrogen stream, i.e., the lower the volumetric flow rate, the higher the relative humidity. The hygrometer HMT 333 was chosen to measure the relative humidity of the outgoing nitrogen stream at five levels. The test data indicated that good agreement was observed between the measured humidity and the observed humidity from the Vaisala, Inc. Moreover, this method offers high sensitivity and fast response. Therefore, the relative humidity of the outgoing nitrogen stream can be used as a secondary humidity reference.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.