Abstract

Abstract The calibration and accuracy of the Eppley precision infrared radiometer (PIR) is examined both theoretically and experimentally. A rederivation of the fundamental energy balance of the PIR indicates that the calibration equation in common use in the geophysical community today contains an erroneous factor of the emissivity of the thermopile. If a realistic value (0.98) for the emissivity is used, then this leads to errors in the total flux of 5–10 W m−2. The basic precision of the instrument is found to be about 1.5% of the total IR irradiance when the thermopile voltage and both dome and case temperatures are measured. If the manufacturer’s optional battery-compensated output is used exclusively, then the uncertainties increase to about 5% of the total (20 W m−2). It is suggested that a modern radiative transfer model combined with radiosonde profiles can be used as a secondary standard to improve the absolute accuracy of PIR data from field programs. Downwelling IR fluxes calculated using the ...

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.