Abstract
The article describes a new interferometric system for measuring the coil displacement in the BIPM Kibble balance. The system has led to a significant reduction in both the type A and type B uncertainty of the Kibble balance measurement. Compared to the previous version, the new instrument doubles the measurement signal thanks to a novel design. The optical and mechanical setup is more compact, more stable, and less sensitive to vibration noise. It is also more easily aligned, particularly for adjusting the interferometer beams vertically.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: IEEE Transactions on Instrumentation and Measurement
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.