Abstract

A modified Rheodyne 7520 microsample injector was used as a new solid phase microextraction (SPME)–liquid chromatography (LC) interface. The modification was focused on the construction of a new sample rotor, which was built by gluing two sample rotors together. The new sample rotor was further reinforced with 3 pieces of stainless steel tubing. The enlarged central flow passage in the new sample rotor was used as a desorption chamber. SPME fiber desorption occurred in static mode. But all desorption solvent in the desorption chamber was injected into LC system with the interface. The analytical performance of the interface was evaluated by SPME–LC analysis of PAHs in water. At least 90% polycyclic aromatic hydrocarbons (PAHs) were desorbed from a polyacrylonitrile (PAN)/C18 bonded fuse silica fiber in 30s. And injection was completed in 20s. About 10–20% total carryovers were found on the fiber and in the interface. The carryover in the interface was eliminated by flushing the desorption chamber with acetonitrile at 1mLmin−1 for 2min. The repeatability of the method was from 2% to 8%. The limit of detection (LOD) was in the mid pgmL−1 range. The linear ranges were from 0.1 to 100ngmL−1. The new SPME–LC interface was reliable for coupling SPME with LC for both qualitative and quantitative analysis.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.