Abstract

A new hybrid (numerical-physical) simulator of the respiratory system, designed to simulate spontaneous and artificial/assisted ventilation of preterm and full-term infants underwent preliminary evaluation. A numerical, seven-compartmental model of the respiratory system mechanics allows the operator to simulate global and peripheral obstruction and restriction of the lungs. The physical part of the simulator is a piston-based construction of impedance transformer. LabVIEW real-time software coordinates the work of both parts of the simulator and its interaction with a ventilator. Using clinical data, five groups of "artificial infants" were examined: healthy full-term infants, very low-birth-weight preterm infants successfully (VLBW) and unsuccessfully extubated (VLBWun) and extremely low-birth-weight preterm infants without (ELBW) and with bronchopulmonary dysplasia (ELBW_BPD). Pressure-controlled ventilation was simulated to measure peak inspiratory pressure, mean airway pressure, total (patient+endotracheal tube) airway resistance (R), total dynamic compliance of the respiratory system (C), and total work of breathing by the ventilator (WOB). The differences between simulation and clinical parameters were not significant. High correlation coefficients between both types of data were obtained for R, C, and WOB (γ R =0.99, P<0.0005; γ C =0.85, P<0.005; γWOB=0.96, P<0.05, respectively). Thus, the simulator accurately reproduces infant respiratory system mechanics.