Abstract

Abstract The authors describe a high-resolution long-path differential optical absorption spectrometer developed for the measurement of tropospheric hydroxyl radical concentrations. The instrument uses an atmospheric absorption light path of up to 3000-m total length that is folded into an optical multiple-reflection cell of 20-m base length. A frequency-doubled picosecond laser system serves as a light source, which has a large spectral bandwidth of 0.41 nm (FWHM) centered around 308 nm. In combination with a high-resolution Echelle spectrograph and an optical multichannel detector, a spectral range of 0.26 nm which comprises six OH absorption lines of the A2∑+, (ν′ = 0)← X2Π, (ν″=0) transition, can be simultaneously recorded. This large spectral detection range facilitates the subtraction of interfering absorption signals of other atmospheric trace gases (SO2, HCHO, naphthalene) and significantly improves the detection specificity for hydroxyl radicals. A 2σ OH detection limit of 1.5× 106 OH cm−3 for an...

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.