Abstract

Urinary biomarkers are usually reported as a normalized ratio to creatinuria, a lack of correct creatinine determination lead to wrong final data. This study aimed to evaluate the Jaffè method accuracy, proposing a new HPLC-MS/MS method. The validation parameters were: run time of 4 ​min, LOD of 0.02 ​mg/L, CV% lower than 4%, recovery >96.1%, no matrix effect.The comparison between HPLC-MS/MS and Jaffè method showed an average bias, variable from high (1.5–3.00 ​g/L) to low concentrations (below than 0.7 ​g/L) of creatinine, ranging from +31% to −25%. The critical issue for the normalization of biomonitoring data was highlighted.