Abstract
An array of five Si(Li) detectors which partially surrounds a collimated 10 Ci 241Am source, combined with a microprocessor-controlled scanning bed, has been developed for imaging of stable iodine within the thyroid. The sensitivity of the system is sufficient to generate high-quality scan images in a scanning time of approximately 15 minutes, with a spatial resolution of 5.9 mm (FWHM). We have evaluated several phantom studies and used the results to develop a computer code which determines the total thyroid iodine content and thyroid area with acceptable accuracy.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.