Abstract

In this paper, a new calibration apparatus has been developed for gas micro-flow calibration. The apparatus employs both the constant volume method and the constant conductance method. The calibration range from 10−1 to 10−13 Pa·m3/s, and the combined standard uncertainty is 1.2%∼1.8%. When calibrating extremely low gas flow rate, the non-evaporable getter pump is used to reduce the effect of chamber wall outgassing on the measurement results, so that the lower limits of the constant volume method and constant conductance method are extended to 10−10 Pa·m3/s and 10−13 Pa·m3/s, respectively. Then a method for measurement of orifice conductance is proposed, the on-line precise measurement of the orifice conductance is realized; the consistency of flow measurement results by constant volume and constant conductance methods is better than 1.3%, then compared with the same standard leak was calibrated in both constant conductance method of the calibration apparatus and the constant pressure gas micro-flow calibration apparatus of the laboratory, and the deviation of those results was less than 1%.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.