Abstract

Wire meshes commonly used as transmission grids in ion optical configurations has certain unavoidable disadvantages. Apart from the loss in transmission, there are small angle deflections while the ions are passing close to the wires. This is important for high resolution time-of-flight mass spectrometry as this effect impairs both the dynamic range and the resolution. The amount of deflection from various commercially available grids are determined from simulation calculations on representative wire grids. Gridless designs of ion optics are becoming more popular in recent times for sophisticated and sensitive measurements. Here we report such a device for time of flight mass spectrometer that also assists in perpendicular ionization. This ion optics together with a gridless ion reflector gives a high time resolution of more than 40000.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.