Abstract

A new glucose clamp technique for in vivo studies of insulin sensitivity was validated clinically. Eighteen patients (10 males, 8 females, age 35-80 years, body mass index 34.6-17.04) were connected to a computer-assisted artificial pancreas "Betalike R", using a new algorithm based on a "minimal model", to carry out the glucose clamp technique automatically and especially to overcome the well-known problems of its priming phase. We performed the euglycemic hyperinsulinemic clamp in four patients and the hyperglycemic hyperinsulinemic clamp in 14. In one patient both clamps were done. The mean priming time to reach steady-state glycemia was 20 min. Plasma insulin concentrations were measured every 20 min. This new automatic glucose clamp technique enables the priming phase to be run without any significant overshoot, and accidental variations of glycemia in steady state were reduced to a minimum. The system showed satisfactory safety and stability in controlling the patient's glycemia and assured high speed of the priming phase.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.