Abstract
We describe the design improvements incorporated in a new generation of absolute gravimeters, the FG5. A vertically oriented (in-line) interferometer design is used to remove the influence of floor vibration and tilt on the optical path length. The interferometer uses an iodine-stabilized laser as a primary length standard, with circuitry for automatic peak detection and locking. The seismic isolation system is an active long-period seismometer (Super Spring). The new design has improved passive isolation and thermal drift characteristics over previous systems. Programming flexibility and control of the test mass trajectory have been improved. The computer system has also improved real-time analysis and system capability. The FG5 instrument has a higher level of robustness, reliability and ease of use. These design advances have led to an instrumental uncertainty estimate of 1,1 × 10-8 m s-2 (1,0 μGal). Instrument agreement among nine similar devices is 1,8 μGal and observations under optimal conditions exhibit standard deviations of 5 μGal to 8 μGal.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.