Abstract

The extrapolation method may be used to correct X-ray absorption without knowledge of thickness in quantitative microanalysis with the analytical electron microscope. This paper presents a new form of the method which permits the analysis of a region equivalent to a focused beam size or of a specimen with a uniform thickness. The development of the method is based on tilting the specimen and introducing a geometrical factor into the earlier form. The new form is general because it includes not only the data for different thicknesses as used in the earlier form but also the data at different tilt angles. The applicability of the new form is demonstrated with reference to experimental data for a Ni-Al-Ta alloy and of Maher et al. [in: Analytical Electron Microscopy - 1981, p. 29].

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.