Abstract

The main reasons for fine needle aspiration (FNA) failure are insufficient number of cells in the sample, nonrepresentative samples, and contamination of samples. The aim of this study was to measure the number of cells harvested by a new needle system with an aeration aperture (Thyrosampler) in comparison with a conventional FNA system (C-FNA). Under a double-blind setting, 30 aspirations, 15 with each system (C-FNA, Thyrosampler), were done in randomized order and recorded. The median total number of cells was 59 680 cells/mL with C-FNA and 396 400 cells/mL with Thyrosampler. The needle system with the aeration aperture led to a significantly higher cell amount (564% more cells than the conventional system; P < .005) in needle aspiration biopsy. The new system with the vacuum release feature leads to a significantly higher cell amount in needle aspiration biopsy, which is a well-defined benefit.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.