Abstract

In this contribution we present a new evaluation technique used for flow injection analysis (FIA) measurements called projective reference evaluation. Using different FIA systems employing an enzyme cartridge in combination with an oxygen electrode for glucose measurement, an enzyme field effect transistor for urea measurement as well as an enzyme optode for penicillin measurement it will be demonstrated, that the new evaluation technique is able to calculate reliable analyte concentrations even from heavily faulty signals. The occurrence of a fault can be detected very easily, using the new technique. Furthermore, it will be shown, that the linear measurement range of a FIA system can be extended significantly. Applying this evaluation technique the whole information of the analysis system inherently hidden in the measurement signal is exploited. The simplicity of the procedure implied its use for on-line application in a process FIA where a reliable evaluation as well as a fast fault detection is imperative.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.