Abstract

To retrospectively correct for geometrical distortions, a new dynamic field mapping method suitable for dynamic single-shot gradient-echo type echo-planar imaging (GRE-EPI) is proposed. The method requires a single volume additional acquisition and allows the extraction of a field map from each phase volume, assuming invariance across time of the echo time-independent phase component. Performances of the method are assessed using three sets of experiments: the first tests the prerequisite and the modeling; the second tests the method with time-dependent geometrical distortions; and the third presents a comparison with two other methods. Our results legitimize the modeling procedure and demonstrate that the dynamic method is less sensitive to noise than the other methods. A theoretical explanation for this is proposed in the discussion section. Given the minor increase in the acquisition time, this method is well suited for functional magnetic resonance imaging; prospective direction.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.