Abstract The design, construction and performance of a double focusing electron spectrometer are described. The spectrometer has a mean radius of 30 cm, and a maximum transmission of 5% of 4π, at a resolution of 1.5%. Resolutions of 0.07% have been obtained at a transmission of 0.5% of 4π with a source 0.020 inches wide. Three separate criteria of optimum performance have been studied experimentally; the elimination of aberrations associated with φ z the elimination of aberrations associated with φ r , and an intermediate focusing in which the φ z and φ r aberrations are equal and overlap. The performance of this spectrometer under the first of these three criteria is described in detail.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.