Abstract

For the purpose of obtaining the smaller uncertainties for Hp(3) and Dp lens in 90Sr/90Y beta reference fields, a new dose determination method based on the Monte-Carlo simulation was proposed. The conversion coefficients from the absorbed dose in air, at the reference point of the extrapolation ionisation chamber, Dair, det to Hp(3; α) and the conversion factors from Dair, det to Dp lens(α) were calculated with EGSnrc, respectively, for the irradiation angles from 0° to 60°. Compared with the dose determination method in International Organization for Standardization (ISO) 6980 standard, the uncertainty reductions of 7.7-52.8% for Hp(3; α) and 7.9-55.0% for Dp lens(α) were achieved, respectively. In addition, for the conversion coefficients from the reference absorbed dose DR to Hp(3; α), the calculations were performed for more irradiation conditions, which are not included in the current ISO 6980 standard. For the calculations of the conversion factors from DR to Dp lens(α), the eye and head phantoms with Chinese characteristics were utilised, which makes the conversion factors more suitable for use in China.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.